A qualitative and quantitative assessment of the impact of three processing algorithms with halving of study count statistics in myocardial perfusion imaging: filtered backprojection, maximal likelihood expectation maximisation and ordered subset expectation maximisation with resolution recovery

Modi, B N, Brown, J L E, Kumar, G, Driver, R M, Kelion, A D, Peters, A M and Fowler, J C (2012) A qualitative and quantitative assessment of the impact of three processing algorithms with halving of study count statistics in myocardial perfusion imaging: filtered backprojection, maximal likelihood expectation maximisation and ordered subset expectation maximisation with resolution recovery. Journal of Nuclear Cardiology, 19 (5). pp. 945-957. ISSN 1532-6551

Full text not available from this repository.

Abstract

INTRODUCTION

Ordered subset expectation maximisation with depth-dependent resolution recovery (OSEM-RR) is a processing algorithm reported to improve images with halved tracer activity in myocardial perfusion scintigraphy (MPS) compared to filtered backprojection (FBP) using conventional activities. OSEM-RR has not yet been compared with maximal likelihood expectation maximisation (MLEM).

METHODS

39 patients undergoing MPS and two anthropomorphic phantoms (one with, one without an inferior wall insert) had full-time (FT) and half-time (HT) SPECT datasets acquired simultaneously and processed by FBP, MLEM and OSEM-RR. Two experienced reporters scored images of all clinical studies (n=234) for conspicuity of a perfusion defect, with results being compared using Wilcoxon paired and Kappa tests. A quantitative assessment based on mean segmental pixel counts taken from numbers automatically displayed over the 20 segments of Cedars Sinai Autoquant QPS image were compared using Pearson's correlation and Bland Altman analysis.

RESULTS

A small but consistent superior concurrence between FT and HT datasets for OSEM-RR compared to FBP and MLEM was observed for both qualitative and quantitative analyses. OSEM-RR resulted in better definition of the inferior wall defect on the phantom study.

CONCLUSION

OSEM-RR appears superior to both FBP and MLEM in terms of handling reduced count statistics.

Item Type: Article
Schools and Departments: Brighton and Sussex Medical School > Clinical and Experimental Medicine
Subjects: R Medicine > RC Internal medicine > RC0071 Examination. Diagnosis Including radiography
Depositing User: Ellen Thomas
Date Deposited: 02 Nov 2012 15:45
Last Modified: 02 Nov 2012 15:45
URI: http://srodev.sussex.ac.uk/id/eprint/42033
📧 Request an update